|
Introduction Before taking the photometry equipment to the telescope some preliminary calibration must be done. Normally these need be done just once. While it is suggested to start with a high voltage of -950 VDC, once the signal-to-noise calibration is set, you can experiment with other settings. If you decide on a different high voltage setting, you may wish to repeat the signal-to-noise calibration, however, unless the high voltage has been changed by more than 10 volts, it may not be necessary. Signal -to-Noise Adjustment
What High Voltage Setting to Use Note: The system zero points and dead time will change with a high voltage change. The following is optional and discussed for those wishing to understand their photometry system better. If you wish to determine what the optimum high voltage is you for your photomultiplier tube you will need to perform the following experiment. You will need a light-tight box with a small wheat type bulb or LED to act as a constant light source. See Figure 29 for a suggested arrangement. The photomultiplier tube with socket and voltage divider you plan to use in the final equipment along with a magnetic shield should be used. With no light on in the box, set the tube's high voltage for -1,000 VDC and let the tube "cook" for about 30 minutes. This will stabilize the noise counts. Note: It is important that the ambient temperature be constant as it will affect the sensitivity and noise counts. Slight temperature variations (a few degrees) will not be much of a problem, but large swings (10's of degrees) may have significant impact and should be avoided. Next, adjust the high voltage to your starting voltage, usually the lowest you want to check. Now set the brightness of the lamp or LED to get counts in the tens of thousands. The counts will vary slightly due to system noise, but should average close to each other for a constant high voltage and a 10 second integration. Make a table of high voltage vs. counts. You can also prepare a table that varies the high voltage with no light. This will show the system noise vs. high voltage. From the tables compare sensitivity vs. noise and decide what high voltage you think is best. Once the desired voltage has been determined the power supply should be regulated ±1.0 VDC or better. A digital readout of the high voltage is helpful for keeping an eye on it. Figure 33 shows a typical plot of high voltage (-900 VDC to -1,000 VDC) vs. sensitivity. As can be seen, a change of the 100 VDC can produce nearly a 0.3 magnitude change. |